Abstract
Background: This study evaluated whether fluorescence lifetime imaging (FLIm), coupled with standard diagnostic workups, could enhance primary lesion detection in patients with p16+ head and neck squamous cell carcinoma of the unknown primary (HNSCCUP). Methods: FLIm was integrated into transoral robotic surgery to acquire optical data on six HNSCCUP patients' oropharyngeal tissues. An additional 55-patient FLIm dataset, comprising conventional primary tumors, trained a machine learning classifier; the output predicted the presence and location of HNSCCUP for the six patients. Validation was performed using histopathology. Results: Among the six HNSCCUP patients, p16+ occult primary was surgically identified in three patients, whereas three patients ultimately had no identifiable primary site in the oropharynx. FLIm correctly detected HNSCCUP in all three patients (ROC-AUC: 0.90 ± 0.06), and correctly predicted benign oropharyngeal tissue for the remaining three patients. The mean sensitivity was 95% ± 3.5%, and specificity 89% ± 12.7%. Conclusions: FLIm may be a useful diagnostic adjunct for detecting HNSCCUP.
Author supplied keywords
Cite
CITATION STYLE
Weyers, B. W., Birkeland, A. C., Marsden, M. A., Tam, A., Bec, J., Frusciante, R. P., … Farwell, D. G. (2022). Intraoperative delineation of p16+ oropharyngeal carcinoma of unknown primary origin with fluorescence lifetime imaging: Preliminary report. Head and Neck, 44(8), 1765–1776. https://doi.org/10.1002/hed.27078