Determination of the volumetric mass transfer coefficient (kLa) using the dynamic “gas out–gas in” method: Analysis of errors caused by dissolved oxygen probes

116Citations
Citations of this article
309Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

There are many dynamic methods for measuring the volumetric mass transfer coefficient. The “gas out–gas in” method can directly determine the volumetric mass transfer coefficient in a bioreactor system and provide estimates of the volumetric microbial oxygen uptake rate and the average oxygen saturation concentration at the gas–liquid interface. The errors on these parameters are large if the dissolved oxygen probe response time is not considered. For reliable measurements, deconvolution of the oxygen probe measurements must be made. © 1995 John Wiley & Sons, Inc. Copyright © 1995 John Wiley & Sons, Inc.