Abstract
BACKGROUND: A high-volume low-pressure endotracheal tube (ETT) cuff forms folds along its contact with the trachea, allowing mucus leakage into the lungs. We developed a thin-walled ETT cuff made of Lycra polyurethane. METHODS: In vitro, we tested 6 of each of the new prototype Lycra cuff, the Mallinkrodt Hi-Lo ETT (polyvinyl chloride cuff), and the Kimberly-Clark Microcuff ETT (polyurethane cuff), for leakage, in an acrylic mock trachea (inner diameter 20-mm), with a cuff inflation pressure of 20 cm H2O. We poured 15 mL of methylene-blue colored water into the acrylic tube above the cuff and observed for leakage for 24 hours. RESULTS: The Lycra cuffs had no folds upon inflation in the mock trachea and completely prevented fluid leakage for 24 hours (P
Author supplied keywords
Cite
CITATION STYLE
Kolobow, T., Cressoni, M., Epp, M., Corti, I., Cadringher, P., & Zanella, A. (2011). Comparison of a novel Lycra endotracheal tube cuff to standard polyvinyl chloride cuff and polyurethane cuff for fluid leak prevention. Respiratory Care, 56(8), 1095–1099. https://doi.org/10.4187/respcare.01099