New analytical application of antibody-based biosensor in estimation of thyroid-stimulating hormone in serum

17Citations
Citations of this article
31Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Background: Conventionally, ELISA is used to measure thyroid-stimulating hormone (TSH) for diagnosis of thyroid disease. In this study, an immunosensor-based, kinetic-exclusion analysis (KinExA) was used for TSH estimation. Methodology: A PMMA microbead column coated with TSH antigen is formed inside the flow cell. Samples consisting of mouse anti-TSH monoclonal antibody and TSH antigen complex in solution are passed over the beads and the unbound anti-TSH antibody is captured by the TSH-coated beads, followed by passing fluorescent-labeled antibody over the beads to generate signals for analysis. The limit of detection for the assay was 0.4 mIU l-1 and the precision was acceptable. Conclusion: The developed sensor was advantageous due to the automated nature and its convenience, without compromising the sensitivity for estimation of TSH.

Author supplied keywords

Cite

CITATION STYLE

APA

Wani, T. A., Zargar, S., Wakil, S. M., & Darwish, I. A. (2016). New analytical application of antibody-based biosensor in estimation of thyroid-stimulating hormone in serum. Bioanalysis, 8(7), 625–632. https://doi.org/10.4155/bio-2015-0034