Troubleshooting carry-over in the lc-ms analysis of biomolecules: The case of neuropeptide y

16Citations
Citations of this article
53Readers
Mendeley users who have this article in their library.

Abstract

We describe systematic troubleshooting of the carry-over of neuropeptide Y (NPY) in LC-MS analysis. The objective was to remove candidate parts of the LC-MS system that are responsible for carry over oneby-one. The findings indicate that the carry-over of NPY occurs on the column, particularly in the guard column and at the consumable seals of the sample-needle and high-pressure valves. The methodology demonstrates that it is possible to troubleshoot carry-over in an LC-MS system in a systematic and logical manner.

Cite

CITATION STYLE

APA

Yamagaki, T., & Yamazaki, T. (2020). Troubleshooting carry-over in the lc-ms analysis of biomolecules: The case of neuropeptide y. Mass Spectrometry, 8(2), 1–6. https://doi.org/10.5702/massspectrometry.S0083