We performed computer simulations to improve transmission efficiencies of a dual ion funnel system implemented on an FT-ICR MS. We found that the low m/z range from 50 to 150 could be significantly improved by operating the two ion funnels at different RF amplitudes. These new operational conditions could be applied to analyze metabolome samples, which require high sensitivity in the m/z range from 50 to 1,000.
CITATION STYLE
Baek, S. J., Kim, S. Y., Kim, T., & Kim, H. S. (2013). Simulation study to improve ion transmission efficiency in mass spectrometry with a dual ion funnel ion source. Mass Spectrometry Letters, 4(4), 91–94. https://doi.org/10.5478/MSL.2013.4.4.91