Incremental needle insertion system for force and position sensing

0Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.
Get full text

Abstract

An Incremental Needle Insertion System (INIS) which simultaneously measures the force and position of a needle during insertion was designed and fabricated for use in a tissue deformation study to improve realism in medical simulation. The INIS was tested in a fresh frozen cadaver experiment and the position of the needle was plotted and compared to the expected needle path. It was found that the INIS is sufficiently accurate with an average path deviation of 1.55 mm. In addition, INIS was shown to successfully measure the maximum Central Venous Catheterization needle insertion force which ranged from 3.02 N to 3.73 N.

Author supplied keywords

Cite

CITATION STYLE

APA

Brown, D., Gonzalez-Vargas, J. M., Han, D., Miller, S., & Moore, J. (2020). Incremental needle insertion system for force and position sensing. In Frontiers in Biomedical Devices, BIOMED - 2020 Design of Medical Devices Conference, DMD 2020. American Society of Mechanical Engineers (ASME). https://doi.org/10.1115/DMD2020-9012