A Terahertz Fast-Sweep Optoelectronic Frequency-Domain Spectrometer: Calibration, Performance Tests, and Comparison with TDS and FDS

10Citations
Citations of this article
15Readers
Mendeley users who have this article in their library.

Abstract

We report calibration and performance tests of a terahertz fast-sweep optoelectronic frequency-domain spectrometer designed for industrial applications, aimed at quantifying its performance specifications and demonstrating its suitability for envisaged usage. The frequency scale is calibrated using atmospheric water vapour lines and a silicon wafer etalon; the amplitude linearity is verified using a set of silicon plates. Instrument performance is tested by measuring transmission properties of a variety of representative samples and comparing with a time-domain spectrometer and a frequency-domain spectrometer.

Cite

CITATION STYLE

APA

Kutz, J., Liebermeister, L., Vieweg, N., Wenzel, K., Kohlhaas, R., & Naftaly, M. (2022). A Terahertz Fast-Sweep Optoelectronic Frequency-Domain Spectrometer: Calibration, Performance Tests, and Comparison with TDS and FDS. Applied Sciences (Switzerland), 12(16). https://doi.org/10.3390/app12168257