We report calibration and performance tests of a terahertz fast-sweep optoelectronic frequency-domain spectrometer designed for industrial applications, aimed at quantifying its performance specifications and demonstrating its suitability for envisaged usage. The frequency scale is calibrated using atmospheric water vapour lines and a silicon wafer etalon; the amplitude linearity is verified using a set of silicon plates. Instrument performance is tested by measuring transmission properties of a variety of representative samples and comparing with a time-domain spectrometer and a frequency-domain spectrometer.
CITATION STYLE
Kutz, J., Liebermeister, L., Vieweg, N., Wenzel, K., Kohlhaas, R., & Naftaly, M. (2022). A Terahertz Fast-Sweep Optoelectronic Frequency-Domain Spectrometer: Calibration, Performance Tests, and Comparison with TDS and FDS. Applied Sciences (Switzerland), 12(16). https://doi.org/10.3390/app12168257