Designs and performance of three new microprocessor-controlled knee joints

31Citations
Citations of this article
81Readers
Mendeley users who have this article in their library.
Get full text

Abstract

A crossover design study with a small group of subjects was used to evaluate the performance of three microprocessor-controlled exoprosthetic knee joints (MPKs): C-Leg 4, Plié 3 and Rheo Knee 3. Given that the mechanical designs and control algorithms of the joints determine the user outcome, the influence of these inherent differences on the functional characteristics was investigated in this study. The knee joints were evaluated during level-ground walking at different velocities in a motion analysis laboratory. Additionally, technical analyses using patents, technical documentations and X-ray computed tomography (CT) for each knee joint were performed. The technical analyses showed that only C-Leg 4 and Rheo Knee 3 allow microprocessor-controlled adaptation of the joint resistances for different gait velocities. Furthermore, Plié 3 is not able to provide stance extension damping. The biomechanical results showed that only if a knee joint adapts flexion and extension resistances by the microprocessor all known advantages of MPKs can become apparent. But not all users may benefit from the examined functions: e.g. a good accommodation to fast walking speeds or comfortable stance phase flexion. Hence, a detailed comparison of user demands and performance of the designated knee joint is mandatory to ensure a maximum in user outcome.

Cite

CITATION STYLE

APA

Thiele, J., Schöllig, C., Bellmann, M., & Kraft, M. (2019). Designs and performance of three new microprocessor-controlled knee joints. Biomedizinische Technik, 64(1), 119–126. https://doi.org/10.1515/bmt-2017-0053

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free