Abstract
This study compares a new latex immunoseparation method for the direct determination of plasma low-density lipoprotein cholesterol (LDL-C) with the reference procedure for LDL-C (β-quantification) in a pediatric hyperlipidemic population. The direct LDL-C assay has a mean bias of -98 mg/L in a fasting group (n = 96) of patients (mean triglycerides 1057 ± 720 mg/L) and a bias of + 177 mg/L in a nonfasting group (n = 42, mean triglycerides 4854 ± 5457 mg/L). The mean total analytical error calculated from our data is 13.8%. The direct LDL-C assay and the commonly used Friedewald calculation respectively classified 81% and 84% of fasting patients correctly, according to the cutoffs of 1100 and 1300 mg/L for LDL-C set by the National Cholesterol Education Program for pediatric patients. Of combined fasting and nonfasting patients, 80% were correctly classified by the direct LDL-C assay. Therefore, despite several analytical shortcomings, the direct LDL-C assay may be useful in managing hyperlipidemic children without the need for a fasting specimen.
Author supplied keywords
Cite
CITATION STYLE
Harris, N., Neufeld, E. J., Newburger, J. W., Ticho, B., Baker, A., Ginsburg, G. S., … Rifai, N. (1996). Analytical performance and clinical utility of a direct LDL-cholesterol assay in a hyperlipidemic pediatric population. Clinical Chemistry, 42(8), 1182–1188. https://doi.org/10.1093/clinchem/42.8.1182