Quality specifications for evaluation and comparison of performance among external quality assessment schemes in occupational and environmental laboratory medicine

4Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Quality specifications (QS) are proposed for lead in blood and for aluminium, copper, selenium and zinc in serum as part of the aim to set standards of performance for laboratories so that results can be demonstrated to be fit for the purpose to which they are applied. The QS were established taking account of the analytical state-of-the-art, physiological variations in the concentrations of the analyte and the clinical purpose for which the assay is to be used. A procedure was devised that uses these QS to give equivalence of assessment among external quality assessment schemes (EQAS), thus avoiding conflicting information which has been demonstrated in the past. Advantages of this procedure are: to provide direct comparison of performance of laboratories taking part in different schemes, to provide equivalence of assessment of laboratory performance necessary to establish mutual recognition agreements, and to demonstrate the fitness for purpose of results from participants. © Springer-Verlag 2006.

Cite

CITATION STYLE

APA

Taylor, A., Angerer, J., Arnaud, J., Claeys, F., Jones, R. L., Mazarrasa, O., … Weykamp, C. (2006). Quality specifications for evaluation and comparison of performance among external quality assessment schemes in occupational and environmental laboratory medicine. In Accreditation and Quality Assurance (Vol. 11, pp. 440–445). https://doi.org/10.1007/s00769-006-0118-8

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free