3-D dose verification by cone-beam optical CT scanning of PRESAGE dosimeter

2Citations
Citations of this article
12Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

There is a clear need for an accurate and practical dosimeter that is able to verify 3-D dose distributions from complex radiation treatments. The purpose of this study is to evaluate the dosimetric performance of PRESAGE radiochromic plastic dosimeter in conjunction with a cone-beam optical CT scanner, Vista™, for 3-D dosimetry. The cone-beam optical CT scanner presented in this study can perform optical readout in less than 30 min, which makes same day dose verification for treatment possible. For dosimetric accuracy, a complete 3-D dose verification procedures were performed, including dose response calibration with a 12 MeV electron beam, a 6 MV photon square field irradiation, as well as a 5-field IMRT irradiation. This study shows that 3-D dose verification with fast optical CT scan of PRESAGE dosimeter is feasible. First-generation optical CT scanner can be used as a gold standard for optical readout of dosimeters.