Drug monitoring: Simultaneous analysis of lamotrigine, oxcarbazepine, 10-hydroxycarbazepine, and zonisamide by HPLC-UV and a rapid GC method using a nitrogen-phosphorus detector for levetiracetam

72Citations
Citations of this article
54Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

A high-performance liquid chromatography (HPLC) assay using UV detection is described for the simultaneous measurement of the newer generation anti-epileptic medications lamotrigine, oxcarbazepine (parent drug and active metabolite 10-hydroxycarbazepine), and zonisamide. Detection of all four compounds can be done at 230 nm; however, there is a potential interference with zonisamide in patients on clonazepam therapy. Therefore, the method uses dual wavelength detection: 230 nm for oxcarbazepine and 10-hydroxycarbazepine and 270 nm for lamotrigine and zonisamide. In addition, a simple gas chromatography method using a nitrogen-phosphorus detector is described for the measurement of levetiracetam, another of the recently approved anti-epileptic medications. For both methods, limits of quantitation, linearities, accuracies, and imprecisions cover the therapeutic range for drug monitoring of patients. A wide variety of clinical drugs, including other anti-epileptic drugs, do not interfere with these assays. These procedures would be of special interest to clinical laboratories, particularly due to the limited availability of immunoassays for newer generation anti-epileptic medications and that therapeutic uses of these drugs are expanding beyond epilepsy to other neurologic and psychiatric disorders.

Cite

CITATION STYLE

APA

Greiner-Sosanko, E., Giannoutsos, S., Lower, D. R., Virji, M. A., & Krasowski, M. D. (2007). Drug monitoring: Simultaneous analysis of lamotrigine, oxcarbazepine, 10-hydroxycarbazepine, and zonisamide by HPLC-UV and a rapid GC method using a nitrogen-phosphorus detector for levetiracetam. Journal of Chromatographic Science, 45(9), 616–622. https://doi.org/10.1093/chromsci/45.9.616

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free