AutoLock: A semiautomated system for radiotherapy treatment plan quality control

36Citations
Citations of this article
29Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

A semiautomated system for radiotherapy treatment plan quality control (QC), named AutoLock, is presented. AutoLock is designed to augment treatment plan QC by automatically checking aspects of treatment plans that are well suited to computational evaluation, whilst summarizing more subjective aspects in the form of a checklist. The treatment plan must pass all automated checks and all checklist items must be acknowledged by the planner as correct before the plan is finalized. Thus AutoLock uniquely integrates automated treatment plan QC, an electronic checklist, and plan finalization. In addition to reducing the potential for the propagation of errors, the integration of AutoLock into the plan finalization workflow has improved efficiency at our center. Detailed audit data are presented, demonstrating that the treatment plan QC rejection rate fell by around a third following the clinical introduction of AutoLock.

Cite

CITATION STYLE

APA

Dewhurst, J. M., Lowe, M., Hardy, M. J., Boylan, C. J., Whitehurst, P., & Rowbottom, C. G. (2015). AutoLock: A semiautomated system for radiotherapy treatment plan quality control. Journal of Applied Clinical Medical Physics, 16(3), 339–350. https://doi.org/10.1120/jacmp.v16i3.5396

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free