Kompakt-flexible Aktuatoren aus Formgedächtnislegierung zur Formgebung chirurgischer Instrumente

2Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Conventional flexible actuators for minimally invasive interventions come with complex and bulky actuation infrastructure. We present a proof-of-concept study of a compact flexible actuator with variable shape change featuring shape memory alloy wire loops in active-antagonist configurations for a bending and S-shape. The actuator was fabricated using heat-treated wire loops. The evaluation using optical marker tracking revealed a mean maximum absolute displacement at the distal tip of 19 mm and a force of 90 mN reached in under 15 s. For control, a closed-loop regime was evaluated with steady-state errors below 2°.

Cite

CITATION STYLE

APA

Fischer, N., & Mathis-Ullrich, F. (2023). Kompakt-flexible Aktuatoren aus Formgedächtnislegierung zur Formgebung chirurgischer Instrumente. At-Automatisierungstechnik, 71(7), 547–553. https://doi.org/10.1515/auto-2023-0049