A high-speed liquid-chromatographic method for measuring urinary porphyrins

28Citations
Citations of this article
11Readers
Mendeley users who have this article in their library.
Get full text

Abstract

We describe a rapid quantitative and qualitative 'high-performance' liquid-chromatographic (HPLC) method for measuring porphyrins in urine. Direct injection of acidified, filtered urine onto a 3-μm (particle size) 3-cm-long reversed-phase column fully resolves uroporphyrin, hepta-, hexa-, and pentacarboxylic acid porphyrins, and coproporphyrin. Instrument response is linearly related to concentration over the range 25 to 300 nmol/L. The method provides data essential for the differential diagnosis of porphyric states, including porphyria variegata and porphyria cutanea tarda. This relatively inexpensive method requires a run time of only 8 min per sample, making it particularly suitable for routine use in the clinical chemistry laboratory.

Cite

CITATION STYLE

APA

Johnson, P. M., Perkins, S. L., & Kennedy, S. W. (1988). A high-speed liquid-chromatographic method for measuring urinary porphyrins. Clinical Chemistry, 34(1), 103–105. https://doi.org/10.1093/clinchem/34.1.103