Abstract
A rapid and high-throughput fluorescence detection method for zearalenone (ZEN) based on a CuO nanoparticle (NP)-assisted signal amplification immunosensor was developed using an automated sample pretreatment and signal conversion system. CuO NPs with high stability and biocompatibility were used as carriers to immobilize anti-ZEN antibodies. The obtained CuO NP-anti-ZEN can maintain the ability to recognize target toxins and act as both a signal source and carrier to achieve signal conversion using automated equipment. In this process, target toxin detection is indirectly transformed to Cu2+ detection because of the large number of Cu2+ ions released from CuO NPs under acidic conditions. Finally, a simple and high-throughput fluorescence assay based on a fluorescent tripeptide molecule was employed to detect Cu2+, using a multifunctional microporous plate detector. A good linear relationship was observed between the fluorescence signal and the logarithm of ZEN concentration in the range of 16.0–1600.0 µg/kg. Additionally, excellent accuracy with a high recovery yield of 99.2–104.9% was obtained, which was concordant with the results obtained from LC-MS/MS of naturally contaminated samples. The CuO NP-based assay is a powerful and efficient screening tool for ZEN detection and can easily be modified to detect other mycotoxins.
Author supplied keywords
Cite
CITATION STYLE
Xuan, Z., Wu, Y., Liu, H., Li, L., Ye, J., & Wang, S. (2021). Copper oxide nanoparticle-based immunosensor for zearalenone analysis by combining automated sample pre-processing and high-throughput terminal detection. Sensors, 21(19). https://doi.org/10.3390/s21196538