Based on a modified dice-and-fill technique, a PIN-PMN-PT single crystal 1-3 composite with the kerf of 12 μmand pitch of 50 μmwas prepared. The as-made piezoelectric composite material behaved with high piezoelectric constant (d33 = 1500 pC/N), high electromechanical coefficient (kt = 0.81), and low acoustic impedance (16.2 Mrayls). Using lithography and flexible circuit method, a 48-element phased array was successfully fabricated from such a piezoelectric composite. The array element was measured to have a central frequency of 20 MHz and a fractional bandwidth of approximately 77% at-6 dB. Of particular significance was that this PIN-PMN-PT single crystal 1-3 composite-based phased array exhibits a superior insertion loss compared with PMN-PT single crystal and PZT-5H-based 20 MHz phased arrays. The focusing and steering capabilities of the obtained phased array were demonstrated theoretically and experimentally. These promising results indicate that the PIN-PMN-PT single crystal 1-3 composite-based high frequency phased array is a good candidate for ultrasound imaging applications.
Author supplied keywords
Cite
CITATION STYLE
Zhou, W., Zhang, T., Ou-Yang, J., Yang, X., Wu, D., & Zhu, B. (2020). PIN-PMN-PT single crystal 1-3 composite-based 20 MHz ultrasound phased array. Micromachines, 11(5). https://doi.org/10.3390/mi11050524