High Sensitivity and Precision High-Temperature Reversed-Phase LC Analysis of Bevacizumab for Intact Bioanalysis of Therapeutic Monoclonal Antibodies

  • YAMADA T
  • MIZUNO H
  • Zhe MIN J
  • et al.
N/ACitations
Citations of this article
13Readers
Mendeley users who have this article in their library.

Abstract

We optimized several analytical conditions for more sensitive and precise HT-RPLC analysis of the therapeutic monoclonal antibody (mAb), bevacizumab. Specifically, we (1) optimized the sample preparation process to reduce adsorption and aggregation of bevacizumab, (2) introduced a sample concentration process using a centrifugal ultrafiltration unit to increase detection sensitivity, and (3) used another therapeutic mAb as an internal standard to improve analytical precision. The optimized method for bevacizumab analysis was shown to have low detection and quantification limits of 0.010 and 0.032 µg/mL, respectively, good correlation coefficients (r 2 > 0.9997), and good intra-and inter-day precisions within < 12.0 %. This study provides an important methodology for the intact bioanalysis of therapeutic mAbs, not merely their LC measurement.

Cite

CITATION STYLE

APA

YAMADA, T., MIZUNO, H., Zhe MIN, J., TOYO’OKA, T., & TODOROKI, K. (2018). High Sensitivity and Precision High-Temperature Reversed-Phase LC Analysis of Bevacizumab for Intact Bioanalysis of Therapeutic Monoclonal Antibodies. CHROMATOGRAPHY, 39(1), 21–26. https://doi.org/10.15583/jpchrom.2017.014

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free