Abstract
Abstract Accuracy of the acoustic Doppler velocimeter (ADV) is evaluated in this paper. Simultaneous measurements of open-channel flow were undertaken in a 17-m flume using an ADV and a laser Doppler velocimeter. Flow velocity records obtained by both instruments are used for estimating the true (“ground truth”) flow characteristics and the noise variances encountered during the experimental runs. The measured values are compared with estimates of the true flow characteristics and values of variance (〈u′2〉, 〈w′2〉) and covariance (〈u′w′〉) predicted by semiempirical models for open-channel flow. The analysis showed that the ADV sensor can measure mean velocity and Reynolds stress within 1% of the estimated true value. Mean velocities can be obtained at distances less than 1 cm from the boundary, whereas Reynolds stress values obtained at elevations greater than 3 cm above the bottom exhibit a variation that is in agreement with the predictions of the semiempirical models. Closer to the boundary, the measure...
Cite
CITATION STYLE
Voulgaris, G., & Trowbridge, J. H. (1998). Evaluation of the Acoustic Doppler Velocimeter (ADV) for Turbulence Measurements*. Journal of Atmospheric and Oceanic Technology, 15(1), 272–289. https://doi.org/10.1175/1520-0426(1998)015<0272:eotadv>2.0.co;2