Abstract
Triton X-100 (octoxynol 9) is a commercially available surfactant used as a solvent detergent in numerous pharmaceutical applications including virus inactivation. A byproduct formed during its synthesis is 1,4-dioxane, the cyclic dimer of ethylene oxide and a possible carcinogen to humans. The United States Pharmacopoeia (USP) contains a labor-intensive 1,4-dioxane test for Triton X-100. The method couples vacuum distillation to extract the 1,4-dioxane from the Triton X-100 matrix followed by gas chromatography (GC) using a packed column with flame-ionization detecton. In order to provide a more automated and specific test methodology, a headspace GC-mass spectrometry (MS) method has been developed for this application. Analyte quantitation is accomplished by the method of standard additions. The automated sample preparation, coupled with the specificity inherent in high-efficiency capillary column separations together with single-ion MS detection, results in an assay that is more efficient, accurate, and precise than the USP procedure. Performance characteristics of the headspace GC-MS method are contrasted with those characteristics of the USP methodology.
Cite
CITATION STYLE
Poss, M., Couch, T., Odufu, A., McCann, J., Mellon, J., Melnick, B., & Jenke, D. (2003). Determination of 1,4-dioxane impurity levels in Triton X-100 raw material by gas chromatography with mass spectrometric detection. Journal of Chromatographic Science, 41(8), 410–417. https://doi.org/10.1093/chromsci/41.8.410