Does experience influence the forces exerted on maxillary incisors during laryngoscopy? A manikin study using the Macintosh laryngoscope

25Citations
Citations of this article
10Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

The influence of the level of experience of the laryngoscopist on the duration of laryngoscopy, the forces exerted on the tongue and on the maxillary incisors during laryngoscopy, were investigated. Five groups (anaesthetists, residents in anaesthesia, nurse anaesthetists, surgeons and medical students), each consisting of 15 individuals, participated in the study. An intubation manikin was used with a laryngoscope modified so that the forces applied during laryngoscopy could be measured. The mean duration of laryngoscopy in these groups was 23.4 sec, 17.6 sec, 27.1 sec, 26.8 sec and 42.7 sec, respectively. The maximally applied forces on the tongue were 71.7 N, 60.5 N, 65.9 N, 74.2 N and 69.7 N, respectively. The maximally applied forces on the maxillary incisors were 49.9 N, 36.3 N, 41.1 N, 58.3 N and 53.9 N, respectively. These results indicate the level of experience has a significant influence on the duration of laryngoscopy but seems to have little influence on the forces applied to the tongue and the maxillary incisors. © 1995 Canadian Anesthesiologists.

Cite

CITATION STYLE

APA

Bucx, M. J. L., van Geel, R. T. M., Wegener, J. T., Robers, C., & Stijnen, T. (1995). Does experience influence the forces exerted on maxillary incisors during laryngoscopy? A manikin study using the Macintosh laryngoscope. Canadian Journal of Anaesthesia, 42(2), 144–149. https://doi.org/10.1007/BF03028268

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free