Effect of long-detection interval vs standard-detection interval for implantable cardioverter-defibrillators on antitachycardia pacing and shock delivery: The advance III randomized clinical trial

373Citations
Citations of this article
172Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Importance: Using more intervals to detect ventricular tachyarrhythmias has been associated with reducing unnecessary implantable cardioverter- defibrillator (ICD) therapies. Objective: To determine whether using 30 of 40 intervals to detect ventricular arrhythmias (VT) (long detection) during spontaneous fast VT episodes reduces antitachycardia pacing (ATP) and shock delivery more than 18 of 24 intervals (standard detection). Design, Setting, and Participants: Randomized, single-blind, parallel-group trial that enrolled 1902 primary and secondary prevention patients (mean [SD] age, 65 [11] years; 84% men; 75% primary prevention ICD) with ischemic and nonischemic etiology undergoing first ICD implant at 1 of 94 international centers (March 2008-December 2010). Interventions: Patients were randomized 1:1 to programming with long- (n=948) or standard-detection (n=954) intervals. Main Outcomes and Measures: Total number of ATPs and shocks delivered for all episodes (primary outcomes) and inappropriate shocks, mortality, and syncopal rate (secondary outcomes). Results: During a median follow-up of 12 months (interquartile range, 11-13), long-detection group had 346 delivered therapies (42 therapies per 100 person-years, 95% CI, 38-47) vs 557 in the standard-detection group (67 therapies per 100 person-years [95% CI, 62-73]; incident rate ratio [IRR], 0.63 [95% CI, 0.51-0.78]; P < .001). The long-vs the standard-detection group experienced 23 ATPs per 100 person-years (95% CI, 20-27) vs 37 ATPs per 100 person-years (95% CI, 33-41; IRR, 0.58 [95% CI, 0.47-0.72]; P < .001); 19 shocks per 100 person-years (95% CI, 16-22) vs 30 shocks per 100 person-years (95% CI, 26-34; IRR, 0.77 [95% CI, 0.59-1.01]; P =.06), with a significant difference in the probability of therapy occurrence (P < .001); and a reduction in first occurrence of inappropriate shock (5.1 per 100 patient-years [95% CI, 3.7-6.9] vs 11.6 [95% CI, 9.4-14.1]; IRR, 0.55 [95% CI, 0.36-0.85]; P =.008). Mortality (5.5 [95% CI, 4.0-7.2] vs 6.3 [95% CI, 4.8-8.2] per 100 patient-years; HR, 0.87; P =.50) and arrhythmic syncope rates (3.1 [95% CI, 2.6-4.6] vs 1.9 [95% CI, 1.1-3.1] per 100 patientyears; IRR, 1.60 [95% CI, 0.76-3.41]; P =.22) did not differ significantly between groups. Conclusions and Relevance: Among patients receiving an ICD, the use of a longvs standard-detection interval resulted in a lower rate of ATP and shocks, and inappropriate shocks. This programming strategy may be an appropriate alternative. Trial Registration: clinicaltrials.gov Identifier: NCT00617175 ©2013 American Medical Association. All rights reserved.

Cite

CITATION STYLE

APA

Gasparini, M., Proclemer, A., Klersy, C., Kloppe, A., Lunati, M., Ferrer, J. B. M., … Arenal, A. (2013). Effect of long-detection interval vs standard-detection interval for implantable cardioverter-defibrillators on antitachycardia pacing and shock delivery: The advance III randomized clinical trial. JAMA, 309(18), 1903–1911. https://doi.org/10.1001/jama.2013.4598

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free