Bone loss and degradation of bone quality in the human femur after total hip arthroplasty under stress-shielding by titanium-based implant

6Citations
Citations of this article
19Readers
Mendeley users who have this article in their library.

Abstract

The present work was aimed at understanding the stress-shielding caused by hip-joint implantation into a femur by using a human cadaver with a cementless hip implant. In particular, bone quality was assessed from the standpoint of preferential c-axis orientation of biological apatite (BAp). Comparing the implanted side to the non-implanted side, a finite element analysis (FEA) indicated that artificial hip-joint implantation had a significant stress-shielding on the femur. The results also showed a remarka-ble decrease in the degree of preferential BAp orientation as well as bone loss in Haversial canal in the medial-proximal femur. This is the first report showing a reduction in the degree of preferential BAp orientation due to a stress-shielding after artificial hip-joint implantation. Since preferential BAp orientation is an important parameter for determining bone mechanical function, these findings should be taken into account in future artificial hip-joint designs, especially those involving the stem component. © 2012 The Japan Institute of Metals.

Cite

CITATION STYLE

APA

Noyama, Y., Miura, T., Ishimoto, T., Ikeo, N., Niinomi, M., & Nakano, T. (2012). Bone loss and degradation of bone quality in the human femur after total hip arthroplasty under stress-shielding by titanium-based implant. Nippon Kinzoku Gakkaishi/Journal of the Japan Institute of Metals, 76(7), 468–473. https://doi.org/10.2320/jinstmet.76.468

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free