Multiple injection techniques for microfluidic sample handling

46Citations
Citations of this article
23Readers
Mendeley users who have this article in their library.
Get full text

Abstract

This paper presents an experimental and numerical investigation into electrokinetic focusing flow injection for bioanalytical applications on 1 × N (i.e., 1 sample inlet port and N outlet ports) and M × N (i.e., M sample inlet ports and N outlet ports) microfluidic chips. A novel device is presented which integrates two important microfluidic phenomena, namely electrokinetic focusing and valveless flow switching within multiported microchannels. The study proposes a voltage control model which achieves electrokinetic focusing in a prefocusing sample injection system and which allows the volume of the sample to be controlled. Using the developed methods, the study shows how the sample may be prefocused electrokinetically into a narrow stream prior to being injected continuously into specified outlet ports. The microfluidic chips presented within this paper possess an exciting potential for use in a variety of techniques, including high-throughput chemical analysis, cell fusion, fraction collection, fast sample mixing, and many other applications within the micrototalanalysis systems field.

Cite

CITATION STYLE

APA

Fu, L. M., Yang, R. J., Lee, G. B., & Pan, Y. J. (2003). Multiple injection techniques for microfluidic sample handling. Electrophoresis, 24(17), 3026–3032. https://doi.org/10.1002/elps.200305528