Abstract
The results of a study of some ways to improve spectroscopy characteristics of the CdZnTe quasi-hemispherical detectors when working in high gamma radiation fluxes are presented. It was shown that the use of IR illumination with a wavelength of 1050 nm or 1200 nm or at slight warm-up of the detector to +30°C … +40°C can significantly improve spectroscopy performance of the CdZnTe detectors of size 3.5 mm × 3.5 mm × 1.75 mm when operating in a tested gamma-radiation field with a dose rate up to 590 mGy/h.
Cite
CITATION STYLE
Loutchanski, A., Fjodorovs, V., Ivanov, V., & Ogorodniks, V. (2020). Application of CdZnTe Quasi-Hemispherical Detectors in Strong Gamma-Radiation Fields. EPJ Web of Conferences, 225, 07004. https://doi.org/10.1051/epjconf/202022507004