Optical coherence tomography monitoring of angioplasty balloon inflation in a deployment tester

9Citations
Citations of this article
7Readers
Mendeley users who have this article in their library.

Abstract

We present an innovative integration of an intravascular optical coherence tomography probe into a computerized balloon deployment system to monitor the balloon inflation process. The high-resolution intraluminal imaging of the balloon provides a detailed assessment of the balloon quality and, consequently, a technique to improve the balloon manufacturing process. A custom-built swept-source optical coherence tomography system is used for real-time imaging. A semicompliant balloon with a nominal diameter of 4 mm is fabricated for the experiments. Imaging results correspond to balloon deployment in air and inside an artery phantom. A characterization of the balloon diameter, wall thickness, compliance, and elastic modulus is provided, based on image segmentation. Using the images obtained from the probe pullback, a three-dimensional visualization of the inflated balloon is presented. © 2010 Canadian Crown.

Cite

CITATION STYLE

APA

Azarnoush, H., Vergnole, Ś., Bourezak, R., Boulet, B., & Lamouche, G. (2010, August). Optical coherence tomography monitoring of angioplasty balloon inflation in a deployment tester. Review of Scientific Instruments. https://doi.org/10.1063/1.3465556