Abstract
An ELPI was introduced to the TDMA setup for measurement of effective density and particle mass. This allows measurement of particle mass change also in cases when the particles are non-spherical or have voids. In addition to mass change of a particle, the density of transferred matter can be calculated if either unconditioned or conditioned particles are spherical and the bulk (material) density of the core particle is known. The performance of the system was tested by numerical simulations and laboratory experiments. According to the results, the smallest detectable particle mass change is approximately ±17%. The measured density for the condensing species was within 15% of the bulk density value.Particle mass change caused by condensation of semivolatile components of exhaust gas on diesel soot particles was also demonstrated. Copyright © American Association for Aerosol Research.
Cite
CITATION STYLE
Ristimäki, J., & Keskinen, J. (2006). Mass measurement of non-spherical particles: TDMA-ELPI setup and performance tests. Aerosol Science and Technology, 40(11), 997–1001. https://doi.org/10.1080/02786820600910779