Abstract
This paper reports the application of ampholyte-based isoelectric focusing in poly(dimethylsiloxane) (PDMS) using methylcellulose (MC) to reduce electroosmosis and peak drift. Although the characteristics of PDMS make it possible to fabricate microfluidic chips using soft lithography, unstable electroosmotic flow (EOF) and cathodic drift are significant problems when this medium is used. This paper demonstrates that EOF is greatly reduced in PDMS by applying a dynamic coat of MC to the channel walls and that higher concentrations of MC can be used to increase the viscosity of the electrode solutions in order to suppress pH gradient drift and reduce " compression" of the pH gradient. To illustrate the effect of MC on performance, several fluorescent proteins were focused in microchip channels 5 μm deep by 300 μm wide by 2 cm long in 3-10 min using broad-range ampholytes at electric field strengths ranging from 25 to 100 V/cm. © 2005 American Chemical Society.
Cite
CITATION STYLE
Cui, H., Horiuchi, K., Dutta, P., & Ivory, C. F. (2005). Isoelectric focusing in a poly(dimethylsiloxane) microfluidic chip. Analytical Chemistry, 77(5), 1303–1309. https://doi.org/10.1021/ac048915+