Abstract
Introduction Hearing screening allows the identification of individuals with hearing loss. Aim To determine the sensitivity and specificity of a portable hearing screening device in middle-aged and older adults using the manufacturer scoring and a scoring system proposed by the researchers. Methods In this transversal study, participants underwent anamnesis, otoscopy, and hearing screening using portable equipment. After this, a pure tone audiometry was performed, with participants classified into two groups: with and without hearing loss. The sensitivity and specificity of the hearing screening were calculated for the right and left ears using two methods of interpretation: the originalmethod recommended by the manufacturer (criteria 1) and the method proposed by researchers (criteria 2). Results The sample consisted of 55 individuals, 83.6% (n = 46) of whomwere women. Per criteria 1, the sensitivities were 26.3 (right ear) and 21.4% (left ear). The specificity was 100% for both ears. Using criteria 2, the sensitivity was 94.7 (right ear) and 100% (left ear). The specificity was 74.3 (right ear) and 65.9% (left ear). Conclusion This study showed that the criteria proposed by the manufacturer presented low sensitivity in the hearing screening. The criteria proposed by the researchers to achieve a more efficient performance reached high and balanced values for sensitivity and specificity. Copyright © 2014 by Thieme Publicações Ltda, Rio de Janeiro, Brazil.
Author supplied keywords
Cite
CITATION STYLE
Cardoso, C. L., Bós, Â. J. G., Gonçalves, A. K., Olchik, M. R., Flores, L. S., Seimetz, B. M., … Teixeira, A. R. (2014). Sensitivity and specificity of portable hearing screening in middle-aged and older adults. International Archives of Otorhinolaryngology, 18(1), 21–26. https://doi.org/10.1055/s-0033-1358660
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.