Evaluation of two automated low-cost RNA extraction protocols for SARS-CoV-2 detection

19Citations
Citations of this article
59Readers
Mendeley users who have this article in their library.

Abstract

Background: Two automatable in-house protocols for high-troughput RNA extraction from nasopharyngeal swabs for SARS-CoV-2 detection have been evaluated. Methods: One hundred forty one SARS-CoV-2 positive samples were collected during a period of 10- days. In-house protocols were based on extraction with magnetic beads and designed to be used with either the Opentrons OT-2 (OT-2in-house) liquid handling robot or the MagMAXTM Express-96 system (MMin-house). Both protocols were tested in parallel with a commercial kit that uses the MagMAXTM system (MMkit). Nucleic acid extraction efficiencies were calculated from a SARS-CoV-2 DNA positive control. Results: No significant differences were found between both in-house protocols and the commercial kit in their performance to detect positive samples. The MMkit was the most efficient although the MMin-house presented, in average, lower Cts than the other two. In-house protocols allowed to save between 350 and 400 for every 96 extracted samples compared to the commercial kit. Conclusion: The protocols described harness the use of easily available reagents and an open-source liquid handling system and are suitable for SARS-CoV-2 detection in high throughput facilities. Copyright:

Cite

CITATION STYLE

APA

Lázaro-Perona, F., Rodriguez-Antolín, C., Alguacil-Guillén, M., Gutiérrez-Arroyo, A., Mingorance, J., & García-Rodriguez, J. (2021). Evaluation of two automated low-cost RNA extraction protocols for SARS-CoV-2 detection. PLoS ONE, 16(2 February). https://doi.org/10.1371/journal.pone.0246302

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free