We report results of the studies relating to the development of an efficient biosensor for non-invasive detection of CYFRA-21-1 cancer biomarker. We used a low dielectric constant material (nanostructured yttrium oxide, nY2O3) for the fabrication of the biosensing platform. The nY2O3 was synthesized via solvothermal process and functionalized using 3-aminopropyl triethoxy silane (APTES). Electrophoretic deposition (EPD) of the functionalized nanomaterial (APTES/nY2O3) onto an indium tin oxide (ITO)-coated glass electrode was conducted at a DC potential of 50V for 60s. The EDC-NHS chemistry was used for covalent immobilization of -COOH bearing monoclonal anti-CYFRA-21-1 onto -NH2 groups of APTES/nY2O3/ITO electrode. To avoid the non-specific interaction on the anti-CYFRA-21-1/APTES/nY2O3/ITO immunoelectrode, bovine serum albumin (BSA) was used. X-ray diffraction (XRD), transmission electron microscopy (TEM), and field emission scanning electron microscopy (FESEM) were utilized for structural and morphological studies, whereas Fourier-transform infrared spectroscopy (FTIR) was used for the bonding analysis. Cyclic voltammetry (CV) and electrochemical impedance spectroscopy (EIS) techniques were used for electrochemical characterization and response studies of fabricated electrodes. The fabricated immunosensor (BSA/anti-CYFRA-21-1/APTES/nY2O3/ITO) exhibited linearity in the range of 0.01-50 ng·mL-1, sensitivity of 226.0Ω·mL·ng-1, and lower detection limit of 0.01·ng·mL-1. A reasonable correlation was observed between the results obtained using this biosensor and concentration of CYFRA-21-1 measured through ELISA (enzyme-linked immunosorbent assay) technique in salivary samples of oral cancer patients.

Cite

CITATION STYLE

APA

Kumar, S., Panwar, S., Kumar, S., Augustine, S., & Malhotra, B. D. (2019). Biofunctionalized nanostructured yttria modified non-invasive impedometric biosensor for efficient detection of oral cancer. Nanomaterials, 9(9). https://doi.org/10.3390/nano9091190

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free