Influence of RF spoiling on the stability and accuracy of T1 mapping based on spoiled FLASH with varying flip angles

203Citations
Citations of this article
231Readers
Mendeley users who have this article in their library.
Get full text

Abstract

There is increasing interest in quantitative T1 mapping techniques for a variety of applications. Several methods for T1 quantification have been described. The acquisition of two spoiled gradient-echo data sets with different flip angles allows for the calculation of T 1 maps with a high spatial resolution and a relatively short experimental duration. However, the method requires complete spoiling of transverse magnetization. To achieve this goal, RF spoiling has to be applied. In this work it is investigated whether common RF spoiling techniques are sufficiently effective to allow for accurate T1 quantification. It is shown that for most phase increments the apparent T1 can deviate considerably from the true value. Correct results may be achieved with phase increments of 118.2° or 121.8°. However, for these values the method suffers from instabilities. In contrast, stable results are obtained with a phase increment of 50°. An algorithm is presented that allows for the calculation of corrected T1 maps from the apparent values. The method is tested both in phantom experiments and in vivo by acquiring whole-brain T1 maps of the human brain. © 2008 Wiley-Liss, Inc.

Cite

CITATION STYLE

APA

Preibisch, C., & Deichmann, R. (2009). Influence of RF spoiling on the stability and accuracy of T1 mapping based on spoiled FLASH with varying flip angles. Magnetic Resonance in Medicine, 61(1), 125–135. https://doi.org/10.1002/mrm.21776

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free