Fast determination of indoor radon (222Rn) concentration using liquid scintillation counting

12Citations
Citations of this article
17Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

The indoor 222Rn radionuclide was directly absorbed in typical 20 ml glass scintillation vials by passing −3 dm3 of ambient air through 16 ml of water-immiscible non-volataile scintillation cocktail Ultima-Gold F for 10 min. The activity of radon and its two α-emitting daughters: 218Po and 214Po, was determined with the BetaScout low-background liquid scintillation counter. The limit of 222Rn detection is 9 Bq/m3, and the quantification limit with 20% relative accuracy is 28 Bq/m3. The results of the indoor Rn measurement in different houses showed good consistency with results obtained from a Sarad EQF 3220 device.

Cite

CITATION STYLE

APA

Bem, H., Długosz-Lisiecka, M., Janiak, S., Mazurek, D., & Szajerski, P. (2017). Fast determination of indoor radon (222Rn) concentration using liquid scintillation counting. Journal of Radioanalytical and Nuclear Chemistry, 312(2), 337–342. https://doi.org/10.1007/s10967-017-5226-x