Studying the chromatographic separation of caffeine, theophylline, and theobromine

  • Hassan H
  • Salih M
  • Alhassani M
N/ACitations
Citations of this article
6Readers
Mendeley users who have this article in their library.

Abstract

Theobromine, theophylline, and caffeine retention times in a C18 HPLC column (stationary phase) were investigated as a function of mobile phase flow rate, mobile phase composition, and column temperature. When the mobile phase flow rate increased from 0.1 to 1 mL/min and the methanol concentration increased from 5 to 30%, the retention time and peak width of these three compounds were found to be reduced. While there was a small influence of increasing the mobile phase flow rate on the resolution, decreasing the methanol concentration in the mobile phase considerably reduced the resolution. In addition, the mobile phase flow rate and composition were determined to have a more significant impact than the column temperature. According to the findings, theobromine, theophylline, and caffeine were most effectively separated by liquid chromatography at a flow rate of 0.5 mL/min with a mobile phase methanol concentration of 15%.

Cite

CITATION STYLE

APA

Hassan, H. A., Salih, M. H., & Alhassani, M. H. (2025). Studying the chromatographic separation of caffeine, theophylline, and theobromine. Iraqi Journal of Chemical and Petroleum Engineering, 26(2), 143–151. https://doi.org/10.31699/ijcpe.2025.2.15