Polyethylene failure in two total knees: Wear of thin, metal-backed PCA tibial components

27Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

Abstract

Two women underwent a PCA total knee arthroplasty of the metal-backed type in which the tibial insert was only 4 mm thick in its thin-nest portion. After 2 and 4 years, both women had pain on weight bearing due to polyethylene breakdown causing metal-to-metal contact. Such thin polyethylene components should not be used for knee replacements. © 1990 Informa UK Ltd All rights reserved: reproduction in whole or part not permitted.

Cite

CITATION STYLE

APA

Lindstrand, A., Ryd, L., & Stenström, A. (1990). Polyethylene failure in two total knees: Wear of thin, metal-backed PCA tibial components. Acta Orthopaedica, 61(6), 575–577. https://doi.org/10.3109/17453679008993587