Evaluation of breathing interplay effects during VMAT by using 3D gel measurements

18Citations
Citations of this article
44Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Respiratory motion during dynamic radiotherapy may affect the absorbed dose distribution both by dose-reducing smoothing and by more complicated interplay effects. In this study we present a novel method to determine the relative importance of these two effects. For the two dynamic deliveries studied in this work, the expected target dose reduction due to the smoothing effect was estimated by measurements convolved by the motion function. Remaining absorbed dose differences were attributed to interplay effects between the motion of the gel phantom and the movement of the modulating MLC leaves during modulated arc radiotherapy. The total dosimetric effect due to breathing motion and dynamic MLC motion during VMAT delivery resulted in an average of about 4% target dose reduction. Comparing with only the smoothing effect, the average difference was decreased to around 1%, and the remaining distribution was attributed to interplay effects. Although the interplay effects were small compared to the smoothing effect, the standard deviations of 1.4-2.3% (1SD) were larger than the narrow distribution for repeated stationary measurement with a standard deviation between 0.5-0.9% (1SD).

Cite

CITATION STYLE

APA

Ceberg, S., Ceberg, C., Falk, M., Af Rosenschöld, P. M., & Bäck, S. Å. J. (2013). Evaluation of breathing interplay effects during VMAT by using 3D gel measurements. In Journal of Physics: Conference Series (Vol. 444). Institute of Physics Publishing. https://doi.org/10.1088/1742-6596/444/1/012098

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free