Eleven-year results of the ABG I hip replacement

10Citations
Citations of this article
13Readers
Mendeley users who have this article in their library.
Get full text

Abstract

The ABG I uncemented hip prosthesis has demonstrated an unacceptably high failure rate and poor functional outcome in our patients. A prospective review was performed of 38 implanted hips with a mean follow-up of 11 years. Sixteen hips have been revised or are candidates for revision hip surgery, due principally to polyethylene wear and pelvic osteolysis. Contributing factors were likely to be the poor wear characteristics of the polyethylene liners-which were gamma irradiated in air-and the thinness of these liners. Of further concern were the poor pain and function scores in the remaining hips [mean Harris Hip Scores (HHS) of 70 (SD 22) at 11-year follow-up]. The influence of the surgical learning curve on clinical results may have been an important factor. Undersized femoral components were associated with poor pain and functional scores in five patients with HHS < 60. All ABG I hip implants must remain under careful, long-term review. © Springer-Verlag 2006.

Cite

CITATION STYLE

APA

Kim, W. Y., & Muddu, B. N. (2006). Eleven-year results of the ABG I hip replacement. International Orthopaedics, 30(3), 182–184. https://doi.org/10.1007/s00264-006-0075-1