Precision and accuracy of FEV1 measurements from the Vitalograph copd-6 mini-spirometer in a healthy Ugandan population

3Citations
Citations of this article
19Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Objective Evaluate the accuracy and precision of the copd-6 mini-spirometer for FEV1 in a rural Ugandan population. Methods In a cross-sectional study, 171 smallholder farmers performed spirometry with copd-6, and a diagnostic-quality spirometer. Results and discussion The copd-6 underestimated FEV1 at low flows and overestimated FEV1 at high flows. Across all participants, the device slightly overestimated FEV1 by 0.04 [0.02; 0.06] L. Calibration data showed similar patterns. Conclusion The copd-6 could be considered as an affordable tool for research on lung function impairment in resource-constrained settings. However, further validation in a study population with obstructive lung disease is needed.

Cite

CITATION STYLE

APA

Hansen, W. A. H., Schlünssen, V., Jørs, E., Sekabojja, D., Ssempebwa, J. C., Mubeezi, R., … Hansen, H. (2021). Precision and accuracy of FEV1 measurements from the Vitalograph copd-6 mini-spirometer in a healthy Ugandan population. PLoS ONE, 16(6 June 2021). https://doi.org/10.1371/journal.pone.0253319