Abstract
Objective Evaluate the accuracy and precision of the copd-6 mini-spirometer for FEV1 in a rural Ugandan population. Methods In a cross-sectional study, 171 smallholder farmers performed spirometry with copd-6, and a diagnostic-quality spirometer. Results and discussion The copd-6 underestimated FEV1 at low flows and overestimated FEV1 at high flows. Across all participants, the device slightly overestimated FEV1 by 0.04 [0.02; 0.06] L. Calibration data showed similar patterns. Conclusion The copd-6 could be considered as an affordable tool for research on lung function impairment in resource-constrained settings. However, further validation in a study population with obstructive lung disease is needed.
Cite
CITATION STYLE
Hansen, W. A. H., Schlünssen, V., Jørs, E., Sekabojja, D., Ssempebwa, J. C., Mubeezi, R., … Hansen, H. (2021). Precision and accuracy of FEV1 measurements from the Vitalograph copd-6 mini-spirometer in a healthy Ugandan population. PLoS ONE, 16(6 June 2021). https://doi.org/10.1371/journal.pone.0253319
Register to see more suggestions